Hologic lorad selenia mammography Доставка из Европы
Год производства: 2007
General specifications:
MODEL Lorad Selenia
CONTROL MONITOR
Detector type (DETECTOR CHARACTERISTICS) 20 high-brightness flat-panel display
Size, cm (in) (MAXIMUM OBJECT) 50.8 (20)
DIAGNOSTIC READING
Station (DIAGNOSTIC READING) SecurView DX with optional multimodality and MRI CAD components
GENERATOR CHARACTERISTICS (POSITIONING ASSEMBLY) Constant potential, high-frequency inverter type
Kv range (X-RAY GENERATOR) 20in 1 kV increments
mAs RANGE (X-RAY TUBE) 3-400
Ma range (IMAGING SYSTEM) 10Dual focus: 10-100 large focus; 10-30 small focus
Time range, sec (X-RAY TUBE) 0.1-4
AEC DETECTOR (POSITIONING ASSEMBLY) Yes
Parameters controlled (X-RAY TUBE) Auto time, auto kV, auto filter
POSITIONING ASSEMBLY
Collimation (POSITIONING ASSEMBLY) Fully automatic
18 x 24 cm (POSITIONING ASSEMBLY) Large flat panel (19 x 23), etc.Automatic, manual override
24 x 30 cm (POSITIONING ASSEMBLY) Automatic
Motion locking mechanism (POSITIONING ASSEMBLY) Automatic
REVOLUTIONS +195 to -150
Vertical, cm (in) (TUBE MOTION) 67-140 (26.5-55)
Sid, cm (POSITIONING ASSEMBLY) 66
Scale guide (POSITIONING ASSEMBLY) Digital display on both sides of compression device
OPTIONAL ACCESSORIES Needle localization kit, 9 additional compression paddles, accessory cabinet
COMPRESSION MECHANISM (POSITIONING ASSEMBLY) Manual and motorized
Force, newtons (X-RAY TUBE) 175 maximum motorized; 300 maximum manual
RADIATION OUTPUT
Mr/sec @ 28 kvp (RADIATION OUTPUT) >800
DICOM 3.0 COMPATIBLE (POWER NEEDED, VAC) Store, store commit, print, query, retrieve, scheduled workflow (DICOM worklist)
RADIATION PROTECTIVE SHIELD (DIGITAL DETECTION PANEL) Yes
L x w, cm (in) (CONTROL CONSOLE) 189.2 x 81.3 (74.5 x 32)
Thickness (DIMENSIONS, cm (in)) 6 mm tempered glass
MAGNIFICATION 1.8x plastic platform
DATA STORAGE, HARD DISK, GB (RADIATION OUTPUT) 3 RAM
X-RAY TUBE
Anode specifications (X-RAY TUBE) Molybdenum rotating
Heat capacity, hu (X-RAY TUBE) 300000
Heat dissipation rate, hu/min (X-RAY TUBE) 60000
Target/filter combinations (X-RAY TUBE) Mo/Rh
Size of focal spot, mm (X-RAY TUBE) 0.1 small; 0.3 large
Grid ratio (IMAGING SYSTEM) 4:1 HTC grid (8:1 cleanup equivalent)
POWER NEEDED 200, 208, 220, 230, 240 VAC nominal 10%; 50/60Hz; single-phase
Dimensions (hxwxd),cm (in) (ANTECHAMBER) 190 x 65 x 110 (74.8 x 25.6 x 43.3) gantry
WEIGHT, kg (lb) (DISPLAY) 300 (661.5)
DIGITAL DETECTION PANEL
Detector type (DETECTOR CHARACTERISTICS) Amorphous seleniumTFT-based aSe direct-capture technology
Detector size, w x h, cm (DIGITAL DETECTION PANEL) 24 x 29, single plate
Max exposure time (DIGITAL DETECTION PANEL) 4 seconds
Image bit depth (DIGITAL DETECTION PANEL) 14
Pixels,m (DIGITAL DETECTION PANEL) 70
Dqe @ lp/mm (DIGITAL DETECTION PANEL) >/-50% at 7 mR x-ray exposure or higher
Dqe @ 5 lp/mm (DIGITAL DETECTION PANEL) 35%
Mtf @ 5 lp/mm (DIGITAL DETECTION PANEL) 1%
OTHER ATTRIBUTES (Interference compensation) Standard Smart Paddles system allows 1 detector to image all breast sizes; HTC grid automatically retracts for geometric magnification views; FlexMonitor allows screen to be positioned for optimal viewing; prior study reviewing and bidirectional communica
FDA CLEARANCE (Interference compensation) Yes
CE MARK (MDD) (Interference compensation) Yes
MARKETING REGION (Interference compensation) Worldwide
Our specifications
Год производства: 2007
Анатолий г. Москва тел.: 89030991110 e-mail:
Объявление подано: 27.05.2018 Просмотров - 800 Номер: 1004119
|
|
|
|